The pds function performs Piecewise Direct Standardization (PDS), a method
proposed by Wang et al. (1991) to transfer spectra from
one instrument to another. Optionally, the local models can be blended with
a global PLSR model (direct standardization), which can lead to better
transfer performance when the instrumental differences are not purely local.
Arguments
- x1
A matrix or data frame containing spectra acquired with the standard (master) instrument.
- x2
A matrix or data frame containing spectra of the same samples acquired with the instrument to be standardized (slave).
- win
An integer specifying the half size of the moving window used for PLSR. A larger value may improve the transfer but will increase computational time. Default is 5.
- ncomp
An integer specifying the number of components to be used in PLSR. Typically, a small number (e.g., 2-5) is sufficient. Default is 2.
- alpha
A numeric value between 0 and 1 specifying the weight for the global PLSR model. A value of 0 (default) corresponds to the original PDS method, while a value of 1 corresponds to using only the global PLSR model.
Value
A list with two components:
transfer_matrixThe \(p \times p\) transfer matrix \(\textbf{F}\).
interceptA vector of length \(p\) containing the intercepts.
Details
For each wavelength \(j\) of the standard (master) instrument, a PLSR model is fitted between the response of the master instrument at \(j\) and the responses of the slave instrument in the window \([j - win, j + win]\) (truncated at the edges of the spectrum). The regression coefficients are stored in a banded transfer matrix \(\textbf{F}\), and the intercepts in a vector \(\textbf{b}_0\). Spectra measured on the slave instrument are then standardized as $$\textbf{X}_{2,std} = \textbf{X}_2 \textbf{F} + \textbf{1}\textbf{b}_0^T$$
When alpha > 0, a global PLSR model mapping the whole slave spectrum to
the whole master spectrum is also fitted, and the transfer matrix and
intercept are \((1 - \alpha)\) times the local ones plus \(\alpha\)
times the global ones.
References
Wang, Y., Veltkamp, D.J., Kowalski, B.R., (1991). Multivariate instrument standardization. Analytical Chemistry, 63(23):2750-2756.
Bouveresse, E., Massart, D.L., (1996). Improvement of the piecewise direct standardization procedure for the transfer of NIR spectra for multivariate calibration. Chemometrics and Intelligent Laboratory Systems, (32)2:201-213.
Examples
data(forageLIBS)
wl <- suppressWarnings(as.numeric(names(forageLIBS)))
spectra <- forageLIBS[which(wl > 760 & wl < 780)] # the K I resonance lines
# a second instrument, with a shifted and less sensitive response
x1 <- as.matrix(spectra[1:30, ])
x2 <- 0.8 * cbind(x1[, -1], x1[, ncol(x1)]) + 50
res <- pds(x1, x2, win = 3, ncomp = 2)
# spectra of the second instrument transferred to the first one
transferred <- sweep(x2 %*% res$transfer_matrix, 2, res$intercept, "+")
max(abs(transferred - x1)) / max(x1)
#> [1] 0.0313053